DC Voltage:
"60.000 mV to 1000.0 V, 6 ranges, Basic accuracy: ±0.025 % rdg. ±2 dgt."
AC Voltage:
"60.000 mV to 1000.0 V, 6 ranges, Frequency characteristics: 20 Hz - 100 kHz. Basic accuracy 45 - 65 Hz : ±0.2 % rdg. ±25 dgt. (True RMS, crest factor 3)"
DC + AC Voltage:
"6.0000 V to 1000.0 V, 4 ranges, Frequency characteristics: 20 Hz - 100 kHz. Basic accuracy 45 - 65 Hz : ±0.3 % rdg. ±30 dgt. (True RMS, crest factor 3)"
Resistance:
"60.000 ? to 600.0 M?, 8 ranges, (Conductance: 600.00 nS, DT4282 only). Basic accuracy: ±0.03 % rdg. ±2 dgt."
DC Current:
"600.00 ?A to 10.000 A, 6 ranges, Basic accuracy: ±0.05 % rdg. ±5 dgt."
AC Current:
"600.00 ?A to 10.000 A, 6 ranges, Basic accuracy 45 - 65 Hz : ±0.6 % rdg. ±5 dgt. (True RMS, crest factor 3). Frequency characteristics: 20 Hz - 20 kHz (at 600 ?A to 600 mA range)"
AC Current(use with Clamp probes) 40 TO 1 kHz:
N/A
Peak:
"DC V measurement: Signal width 4 msec or more (single), 1 msec or more (repeated). AC V, DC/AC A measurement: Signal width 1 msec or more (single), 250 ?sec or more (repeated)"
C(capacitance):
"1.000 nF to 100.0 mF, 9 ranges, Basic accuracy: ±1.0 % rdg. ±5 dgt."
Continuity check:
"Continuity threshold: 20/50/100/500 ?, Response time: 10 ms or more"
Diode test:
"Open terminal voltage: 4.5 V or less, Testing current 1.2 mA or less, Threshold of forward voltage: 0.15 V to 3 V, seven stages"
Frequency:
"AC V, DC+AC V, AC A measurement, at pulse width 1 ?s or more (50 % duty ratio). 99.999 Hz (0.5 Hz or more) to 500.00 kHz, 5 ranges, ±0.005 % rdg. ±3 dgt."
dB conversion:
"Standard impedance setting (dBm), 4 ? to 1200 ?, 20 stages. Display dB conversion value of AC voltage (dBV)"
Temperature(thermocouples):
K: -40.0 °C to 800.0 °C (-40.0 °F to 1472.0 °F). Add accuracy of the Thermocouple probe to main unit accuracy: ±0.5 % rdg. ±3 °C
Other functions:
"Filter function (Remove harmonic noise, use only at 600 VAC, 1000 VAC ranges), Display value hold, Auto hold, Max/Min value display, Sampling select, Relative display, Measurement memory (400 data), Auto-power save, USB communication (option), 4-20 mA % conversion"
Display:
"Main and Sub displays: 5-digits LCD, max. 60000 digits"
Display refresh rates:
"5 times/s (Capacitance measurement: 0.05 to 2 times/s, depending on measured value, Temperature: 1 time/s )"
Power Supply:
"LR6 (AA) alkaline batteries ×4, Continuous use: 100 hours"
Dimensions and mass:
"93 mm (3.66 in) W × 197 mm (7.76 in) H× 53 mm (2.09 in) D, 650 g (22.9 oz) (with test leads holder and batteries)"
Accessories:
"Test Lead L9207-10 ×1, Instruction Manual ×1, LR6 Alkaline Battery ×4"